ENSJ Portable Torque Tester

Contact Now
ENSJ Portable Torque Tester
ENSJ Portable Torque Tester
ENSJ Portable Torque Tester

I. Overview

1.1 Main uses and scope of application

This instrument is a kind of quality control equipment, which can be used for calibration of torque sensors and testing of transmission torque of power system and tightening torque of fasteners such as bolts. It has the characteristics of simple operation, high precision, and easy to carry. It is widely used in various electrical, light industry, machinery manufacturing, scientific research institutions and other industries.

1.2 Product Features

1.2.1 Accuracy: ± 1%.

1.2.2 Can be operated clockwise and counterclockwise.

1.2.3 Peak Hold, MAX and Track modes.

1.2.5 Buzzer and LED warning light.

1.2.6 Three units: N.m, Kgf.cm, lbf.in.

1.2.7 2000 sets of stored values.

1.2.8 Communication transmission function.

1.2.9 You can preset the sleep time to automatically shut down.

1.3 Torque sensor size

1.png

 

mode

lPicture 1

Part

5

50

220

550

2200

A

6.35sp (1/4)

9.5sp (3/8)

12.7sp (1/2)

19sp (3/4)

25.4sp (1)

B

φ38

φ62

φ62

φ94

φ94

C

46

73

73

116

116

D

φ60

φ109

φ109

φ160

φ160

E

6

11

11

26

26

F

φ50

φ90

φ90

φ132

φ132

G

φ4.5

φ8.5

φ8.5

φ12.5

φ12.5

1.4 Variety specifications

model

5

50

220

550

2200

Torque measurement range (N.m)

0.500 ~ 5.00

5.00 ~ 50.00

22.0 ~ 220.0

55.0 ~ 550.0

220 ~ 2200

Precision

± 1%

Data storage capacity

2000 sets

Connector size

1/4

3/8

1/2

3/4

1

Transmission function

Support USB cable

Operation mode

Peak hold (PEAK) / real-time mode / maximum peak

Unit selection

N.m, lbf.in, kgf.cm

Operating temperature

5 ℃ ~ 40 ℃

Storage temperature

-20 ℃ ~ 70 ℃

humidity

90% without condensation

power supply

7.2V Ni-MH battery

Weight (Kg)

0.7

1.4

1.48

4.7

5.0

Overall product structure

Picture 1

2.png3.png

4.png

1.LED warning light 11.USB jack

2.LCD display 12.Power jack

3. Charging indicator 13. Sensor signal input hole

4.Reset button 14.Peak mode * 2

5.Buzzer 15.Preset work item times

6. Power / Cancel button * 1 16. Torque unit

7.Unit selection / setting button 17.Torque display value

8.Down button 18.Maximum

9. Up button 19. Battery level display

10.. Preset job selection button

* 1 Note: When in use, press the\"\" C\"\" key once to turn on the instrument. After turning on, press and hold the\"\" C\"\" key for 2 to 3 seconds to turn off the instrument.

* 2 Note: If\"PEAK\" does not appear on the screen, it means the real-time mode.

Precautions before using the torque tester

2.1 Turn on the power and adjust the torque tester

a. Connect the signal cable between the sensor and the display.

b. Press \"C \" to turn on the power of the tester.

c. After the power is turned on, before using, press \"C \" to reset the tester.

2.2, sensor connection failed

a. If the sensor signal cable is not connected well, the torque tester will display two warning lights on the left and right, and the word\"NOSEN\" will appear on the display.

b. Please confirm that the tester is connected to the sensor, and press \"C \" to restore the normal interface.

2.3 Wake the torque tester in power saving mode

a. In order to save power, the tester enters sleep mode after being left for a period of time (default 20 seconds), the backlight is turned off, and you can wake up the tester by pressing any key; after a period of time (default 2 minutes and 30 seconds), it automatically turns off and stands Time can be set. Press \"C \" again to restart.

2.4 Reset

a. To reset, you need to touch the reset button with a needle, and the instrument will automatically shut down.

b. If the tester does not work properly, reset the tester.

Introduction to menu functions

5.png

3.2 Preset work item settings

This instrument provides ten sets of preset work items from P0 to P9, which can set the operation mode, unit, target value and percentage of the upper and lower limits of the qualified.

6.png

* note1: If the target value is set to full scale (for example, the full scale is 5), pressing the up key when setting the upper limit percentage has no response, and the target value must be between the upper limit and the lower limit.

3.3 Select a preset work item

7.png

3.4 Selecting the torque unit

8.png

3.5, adjust the target torque

9.png

3.6 Operation steps in real-time mode

10.png

* Note1: If the\"\" Er0 \"screen appears, it means that the torque sensor has applied more than 110% of the full-scale torque value, please press \" C \".

* Note2: When the applied torque value reaches 80% of the target torque value, the green LED warning light starts to display, and the buzzer beeps intermittently.

* Note3: When the applied torque value reaches 100% of the target torque value, the red light is on and the buzzer sounds long.

* Note4: When the applied torque value reaches 110% of full scale, the red light is on, the buzzer keeps beeping, and\"Er0 \" appears on the screen. Please stop applying force and press \"C \" to return to zero.

3.7, peak mode operation

11.png

* Note1: If the\"\" Er0 \"screen appears, it means that the torque sensor has applied more than 110% of the full-scale torque value, please press \" C \".

* Note2: If Auto SAVE is set in the menu, the peak value will be saved after 8 seconds.

* Note3: If the storage capacity is full, the first record value will be deleted, the new record value will be stored in the last one, and so on.

* Note4: Can be used with the pass judgment function.

Exceeded the upper limit percentage: It is judged as unqualified, and the red light is on.

Below the lower limit percentage of qualified: judged unqualified, the green light is on.

Within the percentage of the upper and lower limits of the pass, the pass is determined, and the red and green lights are constantly on.

* Note5: When the applied torque value reaches 80% of the target torque value, the green LED warning light starts to display, and the buzzer beeps intermittently.

* Note6: When the applied torque value reaches 100% of the target torque value, the red light is on and the buzzer sounds long.

* Note7: When the value of the applied torque reaches 110% of full scale, the red light is on, the buzzer keeps beeping and \"Er0 \" appears on the screen. Please stop applying force and press \"C \" to return to zero.

3.8 Maximum operating steps

3.8.1. The operation steps are the same as the peak mode.

3.9, peak mode record query

12.png

* Note1: If there is no record value, \"NONE \" is displayed.

Fourth, maintenance and storage

4.1. In order to maintain good accuracy, it is recommended to recalibrate about every year.

4.2. Use over the full scale value may cause damage to the product or loss of accuracy (over 110% of the full scale torque value).

4.3 Do not use the torque tester as a percussion tool.

4.4 Please make sure that the components are connected correctly before use.

4.5 Do not press on the LCD screen.

4.6 Do not use organic solvents to clean the tester, such as alcohol or paint thinner.

4.7. Do not use unauthorized sockets or accessories.

4.8. Do not attempt to disassemble or repair the tester, otherwise the tester may be damaged.

4.9 When the tester is not in use, this product should be stored in an appropriate place.

4.10. Do not touch the power plug with wet hands to avoid the possibility of electric shock.

Five, random attachments

1

Display components

1

2

Sensor assembly

1

3

charger

1

4

Sensor connection line

1

5

Computer cable

1

6

CD

1 serving

7

Manual

1 serving

8

Certificate of conformity

1 serving

9

Warranty Card

1 serving

10

Inspection certificate

1 serving

11

Desiccant

1 package

12

Aluminum alloy portable box key

1

13

Aluminum Carrying Case

1