ED Digital Push Force Meter Instruction Manual

Thank you for purchasing the HF series digital push-pull meter:

The push-pull force meter has the advantages of high accuracy, easy operation, and convenient carrying. It also has various test modes, which are convenient for testers to use, and can store and input test data to the computer for statistical analysis. It can be combined with various test stands and fixtures to form test machines for different purposes. Users can also input the gravity acceleration value of the place of use by themselves to make the test more accurate.

I. Features:

1. High precision and high resolution;

2. Five test modes and three display modes to choose from-to maximize test efficiency;

3. N (Newton), kg (kg) and lb (pound) measurement units are available for selection and conversion.

4. Gravity acceleration setting function—The user can input the precise value of gravity acceleration of the place of use. Make testing and unit conversion more accurate;

5. Peak hold function. Keep the peak display until it is manually cleared;

6, automatic peak function, keep displaying peak automatically after 2 seconds;

7. The upper and lower limits and comparison values ​​can be set for statistical analysis. Buzzer alarm when the comparison value is exceeded;

8, data storage function, can store 128 test values;

9. Data output function, which can input data into the computer through the data line for various analyses;

10.Environmental protection, automatic shutdown in 10 minutes without operation;

11. High-quality rechargeable power supply. The charging voltage is available from 100V to 240V, which can be adapted to most areas at home and abroad. There are also short circuit, leakage, overload protection functions;

12, 2 sets of installation dimensions, suitable for most domestic and foreign test machines, convenient for users to install on the machine;

13. Unique switch contact on-off force test function makes the switch on-off force test more accurate;

14, 6-digit large screen display.

See Table 1 for sensor specifications and Table 2 for push-pull gauge specifications.

1.Sensor specifications

Picture 11.png

2. Model and specification table of external push-pull force sensor

Picture 1

3. Safety precautions:

Precautions

Incorrect operation may damage the instrument or cause serious accidents. This manual indicates important matters to prevent accidents and how to use the instrument. Please read this manual carefully before use, and keep it in a safe place for future reading.

If it is to test the impact load, please choose the model whose maximum load is twice the impact load to be tested;

Cautions

1. During the destructive test, wear a protective mask and gloves to prevent the splashing material from hurting the human body during the test;

2. Do not use fixtures that are damaged or severely deformed. Please refer to the relevant parameters in this manual for self-made jigs (the company has other kinds of jigs, and customers can purchase them as needed);

3. Do not use the instrument beyond its maximum range. Failure to do so may cause damage to the sensor or even an accident;

4. When the test value exceeds 105% of the range, the buzzer will beep continuously. In this case, please quickly remove the added load or reduce the load;

Safety Precautions

1. Please use the matching charger to charge, otherwise it will cause circuit failure or even fire;

2. Do not use a power source other than the rated voltage of the charger, otherwise it may cause electric shock or fire;

3. Do not pull out or insert the plug with wet hands, otherwise it may cause electric shock;

4. Do not pull the power cord of the charger to pull out the plug, so as to avoid the electric wire being torn and getting an electric shock;

5. Please use a soft cloth to clean the machine. Immerse the cloth in water soaked with detergent and wring it dry before removing dust and dirt;

Note: Do not use volatile chemicals to clean the machine (such as volatile agents, thinners, alcohol, etc.)

6. Do not operate the machine in the following environments

(1) Humid environment;

(2) Dusty environment;

(3) Where oil or chemicals are used;

(4) Where there are earthquake sources around.

7. Please use and store within the specified temperature and humidity range, otherwise it may cause instrument failure;

8. Do not disassemble, repair or modify the machine yourself. These actions may cause permanent failure of the instrument;

9. Other outstanding matters needing attention in safety production.

Fourth, the structure name:

1. Structure name diagram

Picture 13.png

Five, screen display instructions

Picture 14.png

PEAK

When \"PEAK \" is displayed, it means Peak mode, and the display shows the peak value until it is manually cleared. When \"AUTO PEAK \" is displayed, it means \"auto peak \" (Peak hold automatic release mode) ), It will be cleared automatically after 2 seconds of displaying the peak value; when \"PEAK \" is not displayed, it means \"track mode \" (real-time load value mode).

2.LO BAT

When the battery voltage drops below 7.0V, \"LO BAT \" appears on the screen, indicating that the voltage is insufficient and needs to be recharged (it can still be tested when charging);

3.MEM

\"MEM \" will be displayed when data is stored. When you press \"DATA \" to view the memory data, MEM \"flashes;

4. In this machine, the thrust (pressure) is displayed by default as a negative value (\"-\") and the pull force is positive (\"+ \" is not displayed)

5.CMP

Compare functions.

This function is activated when the comparison value is set via CPDT (see the corresponding description below). \"CMP \" is displayed. The comparison value defaults to full scale, and can be set separately according to needs when using.

6. In the specifications above HF10K, the last digit of the screen \"Picture 15.png\"Is the letter \" K \", which is combined with \" N \", \" kgf \", \" lbf \"to form \" KN (kiloton) \", \" t (ton) \", \"Klb (thousand pounds) \" as shown:

Picture 16.png

Six, key description (see the figure below)

Picture 17.png

1. \"Startup \" key

When you press this key, the power is turned on, and the\"\" Power\"\" indicator will light up (this indicator will also light up when charging), and the model number will be displayed. After power-on, before pressing the\"\" Zero\"\" key to clear, the zero drift value of the analog signal may be displayed on the screen, just press the\"\" Zero\"\" key to clear it;

2. \"Shutdown \" key

Whenever you press this key, the power is turned off. But the stored memory data will not disappear;

3. \"Memory \" key

When you press this key, the stored test data (memory data) will be called up and displayed on the screen in sequence. At this time, the word\"MEM\" flashes-the number of times of memory is displayed first, and the memory data is automatically displayed after 2 seconds. Press \"Peak \" key to exit the memory;

4. \"Unit \" key

Press this key to switch the measurement unit, and it can display three units of N (Newton), kg (kg) and lb (pound) in a cycle. When the test data is displayed, the unit conversion of the same value can also be completed. In the specifications of HF10K and above, it is composed of the letters\"K\" (KN),\"t (ton)\",\"Klb (kilogram);

5. \"Peak \" key

Each time you press this key, there will be a switch between \"PEAK \" display, \"AUTO PEAK \" display or \"PEAK \" disappearing—that is, peak hold, peak hold automatic release, load real-time value mode switch;

6. \"Communication \" key

When the button is held down and then the button is pressed, the stored test data is output to the computer through the data cable for communication with the computer. After the communication is completed, press the \"Peak \" key to shut down.

In the setting mode of\"MODE\", press this key to change the flashing number between\"0\" and\"9\".

7. \"Storage \" key

When you press this key, the test data displayed on the screen will be stored. The stored data can be recalled by using the\"memory\" key, and can be cleared by using the\"clear\" key, and can also be output to the computer for analysis or printing by using the\"communication\" key.

When test data is stored in the machine, \"MEM \" is displayed;

8. \"Clear \" key

When you press this key, the test value on the screen will be reset to zero.

※ When the weight of the fixture used exceeds 20% of the range or the machine has a load exceeding 20% ​​of the range, it cannot be cleared by pressing the zero key. At this time, a lighter fixture must be used or the applied load should be lifted and cleared zero;

※ Press and hold this key for 4 seconds, all the stored test data can be cleared (it may not be cleared in some states, then restart the machine after powering off and then execute this function to clear all the memory data).

9. \"Settings \" key

For detailed settings, please refer to the description of various setting methods in\"\" Function Settings\"\".

Seven, MODE function settings:

Press the\"\" Settings\"\" key in the power-on state. At this time, the setting items will be displayed. The initial display is\"\" TEST \"(Picture 18.png), The number setting window will appear automatically after 2 seconds. Use the communication key and the\"Save\" key to set the test mode; then press the\"Setting\" key again, and the\"\" AODT \"(Picture 19.png), \"LODT \" (Picture 20.png), \"HIDT \" (Picture 21.png), \"CPDT \" (Picture 22.png), The digital setting windows are automatically displayed 2 seconds after their respective items appear. Use the\"\" Communication\"\" and\"Save\" keys to set the required parameters. \"Setend \" is displayed, indicating that the setting is over and entering the test state.

TEST is represented by four digits, and its specific definition is as follows:

0000 Load real-time value mode 0001 Standard test mode

0002 Push-pull peak mode 0003 Pull-pull peak mode

0004 Thrust peak mode 0005 External node off mode

0006 External node on-off mode

0000 Load real-time value (random tracking) mode

Tracking the change of the test load value in any state shows that when the test load value disappears, it returns to zero. At this time, the PEAK function does not work.

0001 Standard test mode (factory default mode)

Three modes can be set in this mode, namely the real-time load status, peak hold status and automatic peak status. When \"PEAK \" is not displayed on the display, it is the state of the real-time load, and the test value changes with the change of the load. When the\"Peak\" key is displayed, the word\"PEAK \" is displayed. The displayed test value is the maximum value in the test (regardless of the tension and pressure), which needs to be manually cleared; press the\"\" Peak\"\" key again, and the\"AUTO PEAK\" display will show the automatic peak status. The displayed test The value is the maximum value in the test (regardless of tension and pressure), and it will automatically disappear and return to zero after holding the display for 2 seconds.

0002 Push-pull peak mode-the function of simultaneously grasping the maximum load value in both pressure and tension directions. During the connector test, it grasps the function of the maximum load (Fc.Ft) of the positive and negative sides of the insertion force and the extraction force.

※ When the load applied by both the positive and negative sides is greater than the induced value Fa and the load is lower than the induced value Fa, the test is counted once (one cycle) and the test is completed (end). Be sure to set the sensing value Fa (sensing range).

0003 Pull force peak mode-When plugging and unplugging test, only grasp the function of the maximum load (Ft) of pulling force.

0004 Pressure peak mode-During the plug-in test, only the function of the maximum load (Fc) of the plug-in pressure is grasped.

0005 0006 Switch contact on-off force test mode-accurately measure the load value at the moment of contact on-off.

0005 Maximum force value of the external node from the moment of opening to the moment of opening

Connect the two contacts under test to pins 4 and 5 of the data interface (using the data plug in the accessory), press the\"Peak\" key to select the peak mode, and apply pressure to the switch by pushing and pulling the force gauge until The switch is on. The measured force value is the force value required to switch on the switch.

0006 The maximum value of the external node from the moment of being connected to the moment of being disconnected

Connect the two contacts under test to pins 4 and 5 of the data interface (using the data plug in the accessory), press the\"Peak\" key to select the peak mode, and apply pressure to the switch by pushing and pulling the force gauge until The switch is turned off, and the measured force value is the force value required to open the switch.

Test the operating load of switches, buttons, etc. as shown below:

Set in \"PEAK \" state, Fp is the test data value.

After the applied load is greater than the set value of Fa, the load value of the display will stop changing until the contact changes from off to on or from on to off. The displayed value is the test value.

※ Please connect PIN4 and PIN5 of the data interface as the contact signal (use the data plug in the accessory).

※ The test cannot be performed without setting Fa (sensing range).

Picture 23.png

AODT sensor setting

When pushing and pulling at the same time, set the sensing value Fa. For example: thrust test, if the value exceeds the induction value, it means \"push \" to start the test. After that, the value is lower than the induction value, which indicates that the thrust test is over. On the contrary, the reverse direction tensile test can be performed. Pull\"\" to start the test, after which the value is lower than the induction value, indicating that the pull test is over.

LODT test value lower limit setting

Set the lower limit of the test value. If it is lower than the lower limit, it is out of range. \"MIN \" is displayed.

HIDT test value upper limit setting

Set the upper limit of the test value. Above the upper limit is out of range, \"MAX \" is displayed.

CPDT comparison value setting

When the test value exceeds the set comparison value, the buzzer will alarm.

The combination of LODT HIDT CPDT enables the instrument to analyze and judge the test data. When the user does not know the debugging, (preferably in the state of powering off and on again), press and hold the setting button for more than four seconds, and release the hand after hearing the sound of\"\" 嘀\", the instrument can return to the factory default state, the value is as follows :

TEST: 0001 standard test mode; AODT, LODT: 1% of full scale;

HIDT, CPDT: full scale.

Testing

Press the \"On \" key to turn on the power and use the factory default settings to directly perform the test as required or press the Set key to select the test mode and perform the test.

1. Select the appropriate test fixture to be installed on the push-pull gauge (for the self-made fixture, please refer to the relevant data in \"Outline and Installation Dimension Drawing \".

2. Please hold the push-pull meter firmly or install the push-pull meter on a suitable test machine for testing. When testing, please make the test force and the push-pull rod of the push-pull meter in a straight line so as to measure the accurate load .

3. After the test is completed, remove the load, turn off the power, remove the fixture, and clean the objects and put them back into the tool.

In the box for next use.

Picture 24.png

Nine, storage (memory) data

1.Stored data

Only in the PEAK state, after the test is completed, pressing the\"\" Save\"\" key will store the data, and the display will show\"\" MEM\"\". The stored memory data can be saved even after the power is turned off. . When using the\"Memory\" key to view the memory data, \"MEM \" flashes, the number of times of memory appears first, and the value of memory automatically appears after 2 seconds. Press the \"Peak \" key to exit the memory and directly enter the test. The data can also be input into a computer for analysis and processing. This machine can store 128 data. When \"O.E \" is displayed, it means that the 129th data can no longer be stored.

2.Memory clear

In the general load display state, press the\"\" Zero\"\" key for more than four seconds, all the memory data will be cleared and the word\"\" MEM \"will disappear.

This machine can be connected to a computer and enter test data into the computer. View and print the number of tests, average, maximum, minimum, and determine whether the test results meet the set requirements.

 

Ten, changes in the value of gravity acceleration

The user of this machine can input the gravity acceleration value of the place of use. Before turning on, press and hold the\"\" Peak\"\" key, and then press the\"On\" key to enter the gravitational acceleration value setting interface. After entering the new gravitational acceleration value, press the\"Setting\" key to shut down the system. It can be used after restarting.

Eleven, correction method

Please use standard weights regularly to confirm whether the display of the push-pull gauge is accurate. If it is inaccurate, it shall be corrected by the measurement professional according to the following methods:

1. In the shutdown state, press and hold the\"\" memory\"\" button and press the power button simultaneously to start (without loading). The display shows\"\" type\", and a default model is displayed after 2 seconds.

2. Press the\"\" Communication\"\" key to change the number until the model of the push-pull gauge appears, press the\"Setting\" key to confirm, and display\"\" sety\"\", and display the 4-digit gravity acceleration value after 2 seconds. Press the\"\" Set\"\" key again to display 6 digits. After the digits are stable, press the\"Set\" key to confirm. \"Noload \" is displayed, and 6 digits are redisplayed after 2 seconds.

Load the full-scale force value (standard weight) corresponding to this model. After the displayed number is stable, press the\"\" Set\"\" key to display\"\" full \". After 2 seconds, display\"\"setend\"\"and turn off automatically in 2 seconds. , Calibration is complete.

12. Other matters needing attention

Picture 25.png

Thirteen, data interface (9PIN)