ENH Torsion Spring Testing Machine

Model
ENH-500A ENH-1000A ENH-2000A ENH-3000A ENH-5000A ENH-1N.M ENH-2N.M ENH-3N.M ENH-5N.M ENH-10N.M ENH-20N.M
Product Name
立式扭转弹簧试验机 扭转弹簧试验机(带打印)
Contact Now
ENH Torsion Spring Testing Machine
ENH Torsion Spring Testing Machine
ENH Torsion Spring Testing Machine
ENH Torsion Spring Testing Machine

Digital Torsion Spring Testing Machine Instruction Manual

First, the function

ENH series digital torsion spring testing machine is an intelligent multifunctional measuring instrument designed and manufactured for testing and detecting various torsion springs. It has the characteristics of simple operation, high precision, full functions, and convenient portability. It is widely used in various electrical, light industry, machinery manufacturing, scientific research institutions and other industries.

Specifications

1. Measurement parameters:

parameter

Picture 1

model

Test torque

Measuring range

Minimum reading

-100A

100 mN﹒ m

1 ~ 100 (mN﹒m)

0.01 mN﹒ m

-200A

200 mN﹒ m

2 ~ 200 (mN﹒m)

0.01 mN﹒ m

-300A

300 mN﹒ m

3 ~ 300 (mN﹒m)

0.01 mN﹒ m

-500A

500 mN﹒ m

5 ~ 500 (mN﹒m)

0.01 mN﹒ m

-1000A

1000 mN﹒ m

10 ~ 1000 (mN﹒m)

0.1 mN﹒ m

-2000A

2000 mN﹒ m

20 ~ 2000 (mN﹒m)

0.1 mN﹒ m

-3000A

3000 mN﹒ m

30 ~ 3000 (mN﹒m)

0.1 mN﹒ m

-5000A

5000 mN﹒ m

50 ~ 5000 (mN﹒m)

0.1 mN﹒ m

-1

1 N﹒ m

0.01 to 1 (N﹒m)

0.0001 N﹒ m

-2

2 N﹒ m

0.02 ~ 2 (N﹒m)

0.0001 N﹒ m

-3

3 N﹒ m

0.03 to 3 (N﹒m)

0.0001 N﹒ m

-5

5 N﹒ m

0.05 to 5 (N﹒m)

0.0001 N﹒ m

-10

10 N﹒ m

0.1 ~ 10 (N﹒m)

0.001 N﹒ m

-20

20 N﹒ m

0.2 ~ 20 (N﹒m)

0.001 N﹒ m

-30

30 N﹒ m

0.3 ~ 30 (N﹒m)

0.001 N﹒ m

-50

50 N﹒ m

0.5 ~ 50 (N﹒m)

0.001 N﹒ m

-100

100 N﹒ m

1 ~ 100 (N﹒m)

0.01 N﹒ m

-200

200 N﹒ m

2 ~ 200 (N﹒m)

0.01 N﹒ m

-300

300 N﹒ m

3 ~ 300 (N﹒m)

0.01 N﹒ m

-500

500 N﹒ m

5 ~ 500 (N﹒m)

0.01 N﹒ m

-1000

1000 N﹒ m

10 ~ 1000 (N﹒m)

0.1 N﹒ m

-2000

2000 N﹒ m

20 ~ 2000 (N﹒m)

0.1 N﹒ m

-3000

3000 N﹒ m

30 ~ 3000 (N﹒m)

0.1 N﹒ m

-5000

5000 N﹒ m

50 ~ 5000 (N﹒m)

0.1 N﹒ m

2. Angle resolution: 0.1 °.

3. Diameter of torque disc: 100mm.

4. Height of tested spring: <80mm.

5. Maximum reading of twist angle: 9999.9 °.

Third, the main features

1. High precision, high resolution, fast sampling speed, large screen display.

2. High precision torque sensor with torque direction display. High-precision rotary encoder, arbitrary angle measurement.

3. The upper and lower limit torque values ​​and the maximum and small rotation angle values ​​can be freely set. When the set range is exceeded, the text flashes and the buzzer alarms.

4. Three parameters display: torque force value, rotation angle, spring stiffness.

5. The three types of units can be converted to each other for selection (N · m, kgf · cm, Ib · in).

6, real-time, peak, automatic peak three modes can be switched at will.

7. The USB interface is used to communicate with the PC. The synchronous test function can be connected to the computer to test. The computer displays the test torque and angle, torque, time, angle and time curve chart, and the detailed measurement parameter records during the test. Do various analyses.

8. Peak hold, automatic release function, free setting time.

9, large storage capacity, can save 99 sets of test data.

10. Automatic shutdown function without operation, time can be set freely.

Product schematic diagram

Picture 1

1.LCD display 2.Set parameter name bar 3.Function buttons

4.Special test head 5.USB interface 6.Power socket

Five: display panel

Picture 1

Six: Button description

1. ON / OFF key: can be used to turn on and off.

2. Angle clear key: In real-time measurement, press this key to correct the angle zero.

3. View key: Use this key to view the stored measurement data during the measurement interface.

4. Zero clear key: During real-time measurement, press this key to correct the torque zero. During the peak and automatic peak, press this key to clear the peak value and return to zero; when viewing the interface, press this key to clear the current stored measurement value, and long press this key to clear all stored measurement values. In the user setting interface, press this key to return to the previous interface without saving data.

5. Peak key: It is used to switch between three measurement modes: real-time, peak and automatic peak in the measurement interface. It has the function of moving up in the setting interface. In the view interface, press this key to view the previous data.

6. Save key: It is used to save the measured data in the measurement interface. It has the function of moving down in the setting interface. In the view interface, press this key to view the next data.

7. Print key: In the measurement interface, press this key to print the stored measurement value (except those without print function). It has the function of moving left in the setting interface.

8. Unit key: Used to switch between N · m, kgf · cm, and Ib · in when measuring interface. It has the function of right shift in the setting interface.

9. Setting key: The user can enter the setting menu through this key on the measurement interface, and press this key to save the data when setting the data.

Seven: User operation instructions

1. Startup display

Startup display welcome and manufacturer information

Picture 1

Display product name, model and version information

Picture 2

2. User main interface display

Picture 3

3. User setting interface

Press the setting key once, as shown below:

Picture 4

Use the up or down arrow to select the setting item. When the setting item is selected to the user, press the setting key to enter the parameter setting interface.

4.Parameter setting interface

In the user setting interface, press the setting key to enter the parameter setting interface, which is displayed as follows:

Setting parameter description: (Taking the maximum torque value as an example)

Picture 1

Use the up and down keys to shift and modify the data. At this time, press the set key to confirm and save the set data (such as pressing the zero key to return without modifying the data). The set data cannot exceed the maximum load value.

(1) Maximum torque value: The user sets the torque upper limit value, which can be set freely as required. When the upper limit value is reached, the number above the parameter name \"torque upper limit value \" flashes, and the buzzer alarms. The upper limit is not higher than full scale.

(2) Minimum torque value: The user sets the torque lower limit value, which can be set freely as required. When the lower limit value is reached, the number above the parameter name \"torque lower limit value \" flashes. The lower limit cannot be higher than the set upper limit.

(3) Setting the minimum storage value: The user sets the minimum storage value according to the storage needs, and data less than this value will not be stored.

(4) Minimum peak hold value setting: The user needs to set freely according to the peak value and the automatic peak measurement. Data smaller than this value is not saved by the peak value.

(5) Maximum deformation angle setting: The user can set the maximum value of the angle, which can be set freely as required. When the upper limit is reached, the number above the parameter name \"Angle Max \" blinks.

(6) Minimum deformation angle setting: The minimum angle can be set by the user, and can be set freely as required.

(7) Automatic peak time setting: The user can freely set the time that the peak value needs to be held from 1 second to 99 seconds according to the automatic peak measurement state.

(8) Automatic shutdown time setting: In the non-operation state, the automatic shutdown time can be freely set from 1 minute to 99 minutes.

(9) Gravity acceleration setting: The user can set the gravity acceleration value according to the location of the area. The default value of this machine is 9.794.

(10) Restore the initial settings: Improper user operations or changes to the data may cause confusion. You can use this setting to restore the data from 1 to 9 to the factory state.

Eight, other function description

1. Description of printing function: print data through micro printer.

2. Communication: communicate with the host computer via USB. The communication protocol is MODBUS-RTU protocol.

3. Save data: Save the data during measurement by the save key.

4. View stored data: View the stored data through the View key.

5. Alarm description: When the measured value exceeds the upper limit, the upper limit indicator will be displayed, and the buzzer will also alarm; when the measured value is lower than the lower limit, the lower limit indicator will be displayed. When the measured value exceeds 120% of the maximum load, it may cause damage to the sensor. When \"Overload \" warning prompt appears, the machine will enter the automatic protection state, you must press the zero key or restart the machine. If you can't measure, please long press the set key, enter the password: up key, down key, and press the set key Restore factory settings interface, select Restore factory settings, and press the set key to restore (if unable to restore, please contact the manufacturer).

Nine, operating environment

1. Operating temperature: -10 ℃ ~ 40 ℃.

2. Operating humidity: ≤90% RH.

3. There is no vibration source and no corrosive environment around.

X. Operating steps

1. Under normal circumstances, plug in the power cord and turn on the power switch. After the display is stable, if the displayed data is not zero, press the \"Clear \" key to clear it to clear the torque value.

2. According to the test needs, press the moment unit conversion button to select the required unit

3. Set up before testing, the specific steps are as parameter setting interface

Twelve Torsion Spring Torque Test Procedures

1. According to the test requirements, select a suitable mandrel, install the mandrel on the mounting hole of the measuring disc, then install the measured spring on the mandrel, and pass the fixed end through the positioning groove and the triangle block in the groove, the pressure block, etc. Fixed and stable.

Picture 1

2. Adjust the fork on the column of the testing machine so that the other end of the torsion spring comes into contact. At the same time, the force value and angle are cleared at the same time.

3. The torque value and angle value displayed on the LCD screen are zero.

4. According to the requirements of torsion spring measurement, rotate the measurement handle smoothly and uniformly, and the technical parameters of the torsion spring to be measured can be displayed.

5. Repeat the above operation to verify the technical parameters of the torsion spring.

Fourteen, matters needing attention, maintenance and repair

1. Don't test the torque by overload. Be sure to test the torque within the test range of the torque tester, otherwise it will damage the instrument and may cause danger.

2. Do not tap the LCD screen and place objects on the LCD screen.

3. Do not press the function keys with nails, sharp objects or pointed objects.

4. Do not use the torque tester in places where water, oil or other liquids are splashed. Store the torque tester in a cool, dry and vibration-free place.

5. Do not loosen the fixing screws of the torque test head.

6. Do not use any power supply other than the rated voltage, otherwise it may cause electric shock or fire.

7. Please do not pull out or insert the plug with wet hands, otherwise it may cause electric shock.

8. Please use a soft cloth to clean the machine, immerse a dry cloth in water soaked with detergent, wring dry, and then remove dust and dirt. Do not use chemicals that emit easily, such as volatile oils, thinners, alcohol, etc.

9. Handle gently during use and handling.

10. Do not disassemble, repair or modify the machine yourself. These actions may cause permanent failure of the instrument.

11. If any trouble occurs, please contact the place of purchase or the company.

12. Within one month from the date of sale of this product, product quality problems occur under normal use and no damage to the appearance. The customer should bring the original sales invoice, valid warranty card and complete packaging to the original place of purchase or the company will replace the same specifications Model products, the replacement products continue the warranty period and terms of the original product.

13. Within one year from the date of sale of this product, under normal use conditions, non-human faults are covered by the warranty (the user disassembles the machine or repairs at other service points by the company without warranty), and the customer has the original sales invoice and the valid warranty Contact the original place of purchase for a free one-year warranty.

14. The warranty terms of this product are only applicable to DNH series products sold in the Chinese market. For products that exceed the replacement period and warranty period, customers can inquire about maintenance issues at the original place of purchase or contact the company, and the company will provide paid repairs. .